中国·金沙(js6666ViP认证)网址登录入口|欢迎您

js6666金沙登录入口-欢迎您

Products

联系我们

Shenzhen Xinzhi Technology Co., Ltd

Address: 705, Building 4, Tongtai Times Center, Fuhai Street, Bao'an District, Shenzhen
Phone: 86-755-2780-8770
Fax: 86-755-2780-9501
Email:xuzb@xzkj.net.cn
Website:/

Medical ordinary sphygmomanometer series
当前位置:首 页 » Products » Medical ordinary sphygmomanometer series » PCBA scheme of wrist sphygmomanometer
PCBA scheme of wrist sphygmomanometer
  • PCBA scheme of wrist sphygmomanometer

PCBA scheme of wrist sphygmomanometer

Oscillographic method, 【 boost 】 measurement; Use a 【 3 】/【 4 】 V DC micro air pump to pressurize, and the 【 3 】/【 4 】 V DC rapid exhaust valve leaks; Semiconductor resistive sensors for pressure detection.

The wrist sphygmomanometer PCBA scheme is used for an electronic sphygmomanometer that is equivalent to an ordinary watch in size and can be worn continuously for a long time like a watch.

At present, our company has completed the plan design of watch type sphygmomanometer, and the plan features are as follows

1. Small size, low power consumption, and truly wearable medical devices

2. Using oscillographic method, the measurement method is step-up measurement, with short measurement time and accurate blood pressure measurement.

3. Equipped with step counting function and alarm clock function.

4. Low power Bluetooth transmission function, which uploads measurement data to the cloud through a mobile app. Ability to remotely stop discriminating measurement results

5. Easy to operate

6. Using a micro booster pump and a specially customized wrist strap

Measurement method of wrist sphygmomanometer PCBA scheme:

Oscillographic method, 【 boost 】 measurement; Use a 【 3 】/【 4 】 V DC micro air pump to pressurize, and the 【 3 】/【 4 】 V DC rapid exhaust valve leaks; Semiconductor resistive sensors for pressure detection.


Power and voltage operating range:

3.0VDC or 2 sections of 1.5V dry battery voltage working range 2.2VDC -3.6VDC;


5VDC or 5VUSB or 3.7V lithium battery voltage operating range 3.6VDC -5.0VDC


Measurement range and accuracy:

Pressure range: [0-295mmHg (0-39.3kPa)], accuracy ± 3mmHg (± 0.4kPa);


Pulse rate: 40-180 beats/minute, with an accuracy of ± 3%.


Sleep power consumption requirements:

Under rated working voltage, shutdown current ≤ 30uA


Key:

1-5 touch or touch buttons for selection


Display mode: color TFT, LED screen, LED digital screen, LCD code break screen optional

Voice function: Chinese, English, and other national voices can be achieved. With volume control and alarm function

Memory function: can be achieved for up to 2 * 240 groups

Communication interface: USB2.0, IIC, UART

相关产品

咨询

QQ

在线客服

服务时间:9:00-18:00

电话

服务咨询热线:

0755-2780-8770

微信

二维码

微信咨询

邮箱

电子邮箱

xuzb@xzkj.net.cn